Gentec-eo QE and XLE4 Manuel d'utilisateur Page 16

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 38
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 15
QE Series Joulemeter Instruction Manual Version 3.0 April 2014
16
Model
QE12SP-S-MT
QE12SP-H-MT
Optical Absorber
MT
Spectral Range
1, 7, 14
0.19 20 µm
Calibrated Spectral Range
(QE-QED Series)
1, 7
0.248 2.5 µm
0.3 2.1 µm
Typical Sensitivity
2, 13
100 V/J
Calibration Uncertainty
2, 7, 8, 10,
13
± 3%
Repeatability
< 0.5 %
Max. Pulse Energy 1.064 µm
0.266 µm
With QED @ 1.064 µm
QED @ 0.266 µm
3, 11
0.70 J
0.10 J
1.6 J
0.25 J
Noise Equivalent Energy (NEE)
(Typ)
2, 10
0.8 µJ
Max. Repetition
Rate
2, 4, 10
MAESTRO,S-LINK, M-LINK: 6000 Hz
INTEGRA: 5200Hz
Typical Rise Time
(0-100%)
2, 10
20 µsec
Max. Pulse Width (Typ)
2, 5, 9, 10
10 µsec
Max. Energy Density
11
500 mJ/cm
2
@ 1.064µm, 7ns, 10 Hz
70 mJ/cm
2
@ 532nm, 7ns, 10 Hz
70 mJ/cm
2
@ 266nm, 7ns, 10 Hz
Max. Energy Density with QED
4 J/cm
2
@1064nm, 7nsec, Single shot
2 J/cm
2
@1064nm, 7nsec, 10Hz
0.35 J/cm
2
@532nm, 7nsec, 10Hz
0.3 J/cm
2
@266nm, 7nsec, 10Hz
Max. Average Power
Detector Alone (QE12SP-S-MT):
With optional Heatsink (QE12SP-H-MT):
11, 12
3 W (7.5 W with QED)
5 W (12.5 W with QED)
Max. Power Density
Detector Alone (QE12SP-S-MT):
With optional Heatsink (QE12SP-H-MT):
With QED :
10 W/cm² @ 3 W
10 W/cm² @ 5 W
600 W/cm²
Dimensions (H x W x D)
Detector Alone (QE12SP-S-MT):
With optional Heatsink (QE12SP-H-MT):
36 x 36 x 14 mm
36 x 36 x 33 mm
Weight
Detector Alone (QE12SP-S-MT):
With Optional Heatsink (QE12SP-H-MT):
87 g
117 g
Aperture Size
QE 12:
QED 12:
12 x 12 mm
9 x 9 mm
Aperture Area Size
QE 12:
QED 12:
1.4 cm
2
0.81 cm
2
Vue de la page 15