4.5.3. Trigger
Figure 4-30 Trigger Button
The “Trigger” button enables the camera to capture images only when an electric signal is sent to the
Beamage-3.0 via the SMA connector. This can synchronize the system’s capture rate with a pulsed laser
source. A SMA to BNC adaptor is provided (202273). The input trigger signal can be from 1.1 volts to 24
volts. The rise edge response time is 300 ns. The pulse width of the trigger signal must be between 300
ns and 230 ms.
Figure 4-31 SMA Connector for Trigger Input
4.5.4. Show/Hide Options
Figure 4-32 Show/Hide Options Button
The Show/Hide Options button will show or hide the Divergence, theRelative Position”, “Camera Lens
Calibrationand Fixed Crosshair panels beside the Data Acquisitiontab. By clicking on Show All’ or
“Hide All, one can show or hide both tabs at the same time. The Start LabVIEW Pipeline button will
open the communication channel between the PC-Beamage-3.0 and the LabVIEW driver. Please refer to
section 7 for more information about this function.
Figure 4-33 Show/Hide Options